80
Hand-held Analyzer Types 2250 and 2270 – User Manual
When a printer has been selected under the Printer Used parameter, then the Print Screen
command in the Main Menu list appears. Use this command to print a screen dump of any
screen on the analyzer.
The printer must either be a mobile thermal printer (from AM Tech or Seiko) or accept the
PCL
®
printer language:
• MPS: Mobile Pro Spectrum – a mobile thermal printer from AM Tech (www.amteq.co.kr)
• PCL: printers accepting PCL printer language
• PCL Inkjet: suitable for Inkjet printers and supports colour printing
• PCL Laser: suitable for laser printers
• Seiko DPU: DPU-S245 or DPU-S445 – mobile thermal printers from Seiko Instruments
(www.seikoinstruments.com)
PCL is the Printer Control Language developed by Hewlett-Packard (see www.hp.com). For
PCL printers, you can use the Top and Left Margin parameters to position the print on the
paper, and use Width and Height parameters to set the size of the print.
8.2 Modem/DynDNS Settings
You can use the analyzer for monitoring in remote places and control it using Measurement
Partner Suite BZ-5503 via a telephone connection – wired or wireless – using suitable
modems or via network connections. See the overview of possibilities in Table 7.1. When
you connect via a modem, or via a network connection, you will see the contents of the
analyzer in the Instrument Task as if you had made the connection via USB. You then have
the same possibilities for transferring data to the Archives and organising data on the
analyzer, as if you were connected via the USB connector.
The analyzer is connected directly to the Internet through a modem. BZ-5503 or an Internet
browser will then be able to connect to the analyzer through TCP/IP using the analyzer’s
(global) IP address. The setup of the modem connected to the analyzer is described in section
8.2.2 – please refer to the BZ-5503 online help for details on how to make a connection.
8.2.1 Connection to the Network using a LAN Modem
Using a 3G modem with LAN interface is a simple and very reliable and powerful way of
remotely connecting to the analyzer.
Set up the modem according to its user manual (usually requires connection to a PC) before
you connect the modem to your analyzer. When ready, connect the modem to the analyzer
using a LAN cable. For hardware versions 1 – 3, you will also need the 10/100 Ethernet CF
Card UL-1016. (See “Ordering Information” on page 216.)
The Modem parameter should be set to Off.
The Airlink Raven XE, Intelligent 3G Ethernet Gateway from Sierra Wireless
(www.sierrawireless.com) has been tested.
Note: The SIM card account should include a data subscription and be usable without a pin code.
Comentarios a estos manuales