194
Hand-held Analyzer Types 2250 and 2270 – User Manual
17.4 Setting up the Analyzer for Building Vibration
Measurements
1) Connect the accelerometer to the rear socket. Brüel & Kjær recommends Accelerometer
Type 8344 for low-level building vibration measurements. If it is your first time to use the
microphone, it needs to be created in the transducer database (see section 5.8) and
calibrated (see section 5.3).
2) Select the Sound Level Meter, Frequency Analyzer, Logging, Logging SLM, Enh. Logging
or Enh. Logging SLM project template.
The project template is displayed at the top of the screen, if it does not display the correct
template tap the Project Template Bar at the top of the screen and select the correct
template from the drop-down menu. (See section 3.3.1 for more details on templates.)
3) Tap the Main Menu icon and select Setup.
4) Select Input > Transducer and select appropriate accelerometer.
5) Select Frequency Settings > Low Frequency Option and select On.
This will enable extra options for the Low Frequency parameter (Frequency Analyzer
module only) and for the frequency weighting parameters (Fig.17.5).
Fig.17.5
Example of additional
parameters with the
Low Frequency Option
parameter set to On
6) Set Low Frequency to Extended.
7) Set Broadband (excl. Peak) to Linear,Wm (simultaneous measurement with the two
frequency weightings Linear and Wm).
8) Set Broadband Peak to Wm.
9) Set Spectrum Weighting parameter to Linear or Wm, as required.