Hand-held Analyzer Types 2250 and 2270 User Manual
Note for Microphone Type 4193 and Type 4964
Microphone Type 4193 is delivered with the Adaptor UC-0211; for Microphone Type 4964 it
is ordered separately.
Adaptor UC-0211 is only used when measurements are made on signals that contain
components at frequencies below 1 Hz that should be included in the measurement. This is
only for the Low Frequency parameter set to Very Low.
It is not recommended to use the adaptor for other signals or settings because it reduces the
dynamic range significantly.