152
Hand-held Analyzer Types 2250 and 2270 – User Manual
14.2.1 Interrupted Noise Method
8) Set the number of decays you want to measure per position – the analyzer automatically
controls the generator, measurement of the decays and averaging the decays.
9) Set the Generator as required – set Generator Type = External if you want to control an
external generator (details in Appendix A) – otherwise leave it at Internal to use the
internal generator.
10) Select Noise Type for the internal generator. Pink noise is typically used.
11) Adjust the level of the internal generator output to match the input of the power amplifier
used by setting Level [re. 1 V].
Note: You can manually turn the generator on and off by tapping on the Loudspeaker
icon in the Status Field.
12) Set the Escape Time allowing you to leave the room before the generator is turned on
during the measurement – see Fig.14.3.
13) Set Build-up Time to allow the excitation noise to reach a steady level before the
measurement starts. 1 s is adequate in most ordinary rooms, but should be increased for
larger halls or reverberation rooms.
14) Select the Sound Source used, to optimise the frequency response of the internal generator
output for either a flat power response or optimum power difference between adjacent
1/1- or 1/3-octave bands – thereby eliminating the need for an equalizer to smooth the
response in most cases. Select a type matching your sound source – select Unknown if you
are using a non Brüel & Kjær sound source, or do not want to make a correction to the
frequency response. Go to step 16.
14.2.2 Impulsive Method
15) Set Trigger Level low enough to be sure the impulse will be triggered, but high enough to
avoid triggering on the background noise. A level between 80 and 100 dB is normally
adequate.
16) If you have selected Automatic Save = Yes, then Trigger Repeat can be set to Yes to
automatically start a new measurement when a measurement has been saved. This allows
you to go to another position and make a new impulse, without the need for controlling the
analyzer between the measurements. Observe how the “traffic light” indicates the status of
the measurement (see Table 4.2 and Fig.14.4) making it easy for you to change position
and generate the impulse synchronised with the measurement procedure. Press the
Start/Pause pushbutton to stop the measurement when the last measurement has been
saved.
14.2.3 Signal Recording
17) Set Recording Control = Automatic if you want to record the signal during the measurement.
The recordings can be played back afterwards for identifying the cause of measurements
differing from each other – or (if Excitation = Impulse) for further analysis of the measured