CHAPTER 2
Assembling your Analyzer
17
4) Insert the male plug of the dual 10-pole adaptor into the top socket of Type 2270 (see item
17 in Fig.2.1) and push gently until it snaps into position.
Mounting the Measurement Microphone on the Microphone Holder and Tripod
Mount the microphone and preamplifier combination onto Microphone Holder UA-1317 and
Small Tripod UA-0801, as follows:
1) Set Small Tripod UA-0801 in the required position, and adjust to the required height.
2) Screw Microphone Holder UA-1317 onto the threaded stud on the tripod ball-joint, and
position the holder as required.
3) Push the microphone and preamplifier combination carefully into the microphone
holder, so that it grips onto the body of the microphone and preamplifier. Ensure that the
microphone cable sits properly in the plastic guide.
Once you have carried out these instructions, you are ready to start measuring, see Chapter 3.
2.5.4 Measuring for Convenience
Measurements that need to be done at short notice, with no particular accuracy can be carried
out by simply holding the analyzer or using any combination of hardware components and
accessories shown in Fig.2.4.
2.5.5 Measuring Using Accelerometers
For vibration measurements you need an accelerometer as the measurement transducer. The
accelerometer is connected to the rear socket of the analyzer.
Please refer to Chapter 3 of FFT Analysis Software User Manual BE 1778 for detailed
information about selecting the right accelerometer and how to mount it correctly.