CHAPTER 18
Specifications
211
ELECTRICAL
Uses internally generated electrical signal combined
with a typed-in value of microphone sensitivity
CALIBRATION HISTORY
Up to 20 of the last calibrations made are listed and
can be viewed on the analyzer
Signal Monitoring
Input signal A-, C- or Z-weighted can be monitored
using an earphone/headphones connected to the
headphone socket
Headphone Signal: Input signal can be monitored
using this socket with headphones/earphones
Gain Adjustment:60 dB to 60 dB
Annotations
VOICE ANNOTATIONS
Voice annotations can be attached to the
Reverberation Time Project, to Sources, to Receivers
and to measurements at each Position
Playback: Playback of voice annotations or signal
recordings can be listened to using
earphone/headphones connected to the headphone
socket
Gain Adjustment:60dB to +60dB
TEXT & IMAGE ANNOTATIONS
Text and image (Type 2270 only) annotations can be
attached to the Reverberation Time Project, to
Sources, to Receivers and to measurements at each
Position
GPS ANNOTATIONS
A text annotation with GPS information can be
attached to the project (Latitude, Longitude, Altitude
and position error). Requires connection to a GPS
receiver
Data Management
Project Template: Defines the display and
measurement setups
Project: Measurement data for all positions defined in
a room are stored with the Project Template
Job: Projects are organised in Jobs
Explorer facilities for easy management of data (copy,
cut, paste, delete, rename, open project, create job,
set default project name)
For specifications and details on documenting results
in Qualifier Type 7830 and Qualifier Light Type 7831,
please refer to Product Data BP 1691
Software Specifications – 2-channel Option BZ-7229
2-channel Option BZ-7229 is enabled with a separate
license. It adds 2-channel functionality to Sound Level
Meter, Frequency Analysis, Logging, Enhanced
Logging and Building Acoustics Software and to Signal
recording option on Type 2270.
Note: 2-channel Enhanced Logging is not available for
hardware versions 1-3.
Please refer to Product Data BP 2190 for Dual-
channel Building Acoustics Software specifications
In general the 2-channel Option adds an extra channel
of measurement data. The two channels can have
input from same type of transducers (e.g., two
microphones) or two different transducers (e.g., one
microphone and one accelerometer)
SELF-GENERATED NOISE LEVEL
Typical values at 23°C for nominal microphone open-
circuit sensitivity:
High Range:
Low Range:
Weighting Microphone Electrical Total
“A” 14.6 dB 28.3 dB 28.5 dB
“B” 13.4 dB 26.9 dB 27.1 dB
“C” 13.5 dB 27.0 dB 27.2 dB
“Z” 5 Hz–20 kHz 15.3 dB 31.2 dB 31.3 dB
“Z” 3 Hz20 kHz 15.3 dB 32.1 dB 32.2 dB
Weighting Microphone Electrical Total
“A” 14.6 dB 12.4 dB 16.6 dB
“B” 13.4 dB 11.5 dB 15.6 dB
“C” 13.5 dB 12.9 dB 16.2 dB
“Z” 5 Hz–20 kHz 15.3 dB 18.3 dB 20.1 dB
“Z” 3 Hz20 kHz 15.3 dB 25.5 dB 25.9 dB