Hand-held Analyzer Types 2250 and 2270 User Manual
Excitation Impulsive
Interrupted Noise
Select Impulsive to excite the room with an
impulse. The analyzer triggers on the impulse,
measures the impulse as a sequence of spectra
with 5 ms intervals, and finally backwardly
integrates the measurements to decay curves.
Select Interrupted Noise to excite the room with
noise, interrupt the noise and measure the decays
– all controlled by the analyzer
Number of Decays 1 to 99 Parameter only available if Excitation = Interrupted
Noise.
Specify the number of decays to be measured
automatically and averaged together per position
Trigger Level 0 to 200 dB Parameter only available if Excitation = Impulsive.
Specify the trigger level for the sound level in any
frequency band – the measurement starts as soon
as this level is exceeded. The measurement has a
pre-trigger of 1 s
Trigger Repeat Off
On
Set to On to start a new measurement
automatically, after saving the previous
measurement (and after fulfilling the trigger
conditions).
Parameter only available if Excitation = Impulsive
and Automatic Save = Yes
Measurement Mode Parallel
Serial
Combined
Parallel measures all bands in the spectrum
simultaneously, Serial measures selected bands,
and Combined first measures Parallel and then
Serial in an automated sequence
Table A.5 (Cont.) Measurement Control parameters – for Reverberation Time Software BZ-7227
Parameter Values Comment