236
Hand-held Analyzer Types 2250 and 2270 – User Manual
Stop Level –100 to 200 dB Stop trigger conditions are fulfilled, when level
crosses Stop Level (in accordance with Start
Slope) for at least Stop Duration seconds – set
Stop Level here
Stop Duration 0 to 420 s Stop trigger conditions are fulfilled, when level
crosses Stop Level (in accordance with Start
Slope) for at least Stop Duration seconds – set
Stop Duration here
Trigger Parameter Microphone:
L
Xeq
L
Yeq
L
Aeq,T1,mov
L
Aeq,T2,mov
L
Ceq,T1,mov
L
Ceq,T2,mov
L
Geq
L
G10max
L
G10min
L
Vpeak
L
XFmax
L
XSmax
L
XImax
L
YFmax
L
YSmax
L
YImax
L
XFmin
L
XSmin
L
XImin
L
YFmin
L
YSmin
L
YImin
L
XF(SPL)
L
XS(SPL)
L
XI(SPL)
L
YF(SPL)
L
YS(SPL)
L
YI(SPL)
(Values continue on
next page)
Select which parameter to monitor for the level
trigger. The parameter is based on 1 second
measurements and checked every second
regardless of the logging period and logged
parameters
X = frequency weightings A or B (controlled by
Setup > Frequency Settings > BB (ex. peak)
parameter).
Y = frequency weightings C or Z (controlled by
Setup > Frequency Settings > BB (ex. peak)
parameter).
V = frequency weightings A, B, C or Z
(controlled by Setup > Frequency
Settings > BB Peak parameter)
Note: The parameter values shown are for
single-channel templates. For 2-channel
templates (Type 2270 only) the values are
available for both Ch.1 and Ch.2
Table A.10 (Cont.) Level Trigger parameters
*
Parameter Values Comment