CHAPTER 11
Logging (Optional Module)
121
Fig.11.2 Relationship between trigger parameters
11) Under the Signal Recording parameters, choose the settings for recording the signal while
making a logging – see Chapter 12.
12) Under Input specify Trigger Input if you want to start the signal recording using an
external trigger signal. See details in Appendix A.
To exit the screen, tap the icon.
11.2 Controlling the Measurement
The measurement is controlled in the same way you would control a normal Sound Level
Meter measurement, using Start/Pause , Continue , Back-erase and Reset
pushbuttons, see Chapter 3 for more details.
If you have set the Charge Injection Calibration parameter to On, then CIC is performed at the
start of the measurement and at the end of the measurement. An Exclude marker is set on the
logging profile at the point where CIC is being performed. The update of the Total parameters
is stopped while CIC is being performed. You can see the results of the CIC under Total
parameters, CIC Result 1 and CIC Result 2.
11.2.1 Annotating a Project
While using the logging software you can annotate measurements using the normal method of
adding annotations to a project, before or after a measurement, or while the measurement is
paused. The annotations can then be viewed by tapping on the Annotations icon or tapping the
Main Menu and selecting Explorer from the list of options. section 3.5.
Start Level = 50 dB
Start Duration
100 dB
90 dB
80 dB
70 dB
60 dB
50 dB
40 dB
30 dB
20 dB
10 dB
0
Time in secs.
Trigger Point
Trigger Point
Pre-recording (BZ 7226 only) Post-recording (BZ 7226 only)
2s
Stop Level = 40 dB
Stop Duration
3s
Trigger Parameter = L
Aeq
Start Slope = Rising
L
Aeq
050208/1