
51
Chapter 5
Calibration
5.1 Introduction
Calibration is an adjustment of your analyzer to measure and display correct values. The
sensitivity of the transducer as well as the response of the electronic circuitry can vary slightly
over time, or could be affected by environmental conditions such as temperature and humidity.
While you are unlikely to ever experience a large drift or change in sensitivity with the
analyzer, it is nevertheless good practice to perform regular calibrations, normally before and
after each set of measurements. Often calibration is required by measurement standards, such
as IEC 61672–1.
If you want your analyzer to remind you when the next Accredited/Traceable calibration is due,
you can set it up on the Reminder tab of the Calibration screen. See section 5.9.
You can check the complete measurement chain during long-term measurement using Charge
Injection Calibration. See section 5.10.
5.2 Acoustic Calibration
Acoustic calibration is the preferred calibration method when measuring with microphones,
particularly when standards and regulations require calibration before a measurement. The
method involves applying an acoustic signal of known magnitude and frequency to the
microphone, which calibrates all the analyzer’s components (microphone, preamplifier and
electrical circuitry).
5.3 Sound Level Calibration
To perform acoustic calibration use Sound Calibrator Type 4231. It provides a stable sound
pressure at 1 kHz and has minimal susceptibility to environmental factors. The procedure itself
is relatively simple, and on the analyzer the procedure is referred to as the Standard
Calibration procedure.
The procedure for performing an acoustic calibration and instructions on how to fit the calibrator
are given in the following section.
Comentarios a estos manuales