APPENDIX A
Setup Parameters
245
A.14 Output Socket Signal
Table A.14 Output Socket Signal parameters
*
*. Not available for Reverberation Time Software BZ-7227.
Parameter Values Comment
Source Microphone:
Off
Input X-weighted
Input C-weighted
Input Z-weighted
L
XF
DC Voltage
Generator
Accelerometer and
Direct:
Off
Input Linear
Output to the Output socket on the connector
panel. Select between Off and the input signal
for monitoring purposes.
The L
AF
setting will output the X-weighted
sound level as a voltage between 0 V and 4 V.
Use the DC Voltage parameter to calibrate the
connected equipment by setting the DC
Output.
Use Generator parameter to output the signal
specified by the Generator settings.
Note 1: If you do not want to output the signal,
then select Off to economise the power.
Note 2: X = frequency weighting A or B. ‘A’
requires that the Broadband (excl. Peak)
parameter is set to AC or AZ. ‘B’ requires that
the Broadband (excl. Peak) parameter is set
to BC or BZ.
Note 3: The Generator setting is only
available for Frequency Analysis templates
DC Output (20 mV/dB) 0.0 to 200.0 dB Source = DC Voltage will output:
0V for DC Output = 0dB and
4V for DC Output = 200 dB
Output Gain Input –60.0 dB to 60.0 dB Output gain of the input signal. Key in a gain
value (0.1 dB resolution) for the input signal.
Use ‘@’ to assign the new value for immediate
response at the output – or use the up/down
navigation keys to increment/decrement the
value in steps of 1 dB.
Note: 0 dB means 1 V output for 1 V input.