Vista de pagina 167
156
Hand-held Analyzer Types 2250 and 2270 – User Manual
The third line of information allows you to:
Select the measured position. The selected position is the one highlighted in the table. The
selected position will be the one displayed when selecting the Spectrum or Decay view
Step forwards or backwards through the positions on all displays, using the and
icons
Get information on the quality of the measurement through a quality indicator (smiley)
displayed as an icon:
: means results should be used with caution
: means results may be suspect or missing
No icon means the analyzer found no measurement quality issues
Tap the smiley to get more detailed information about the quality indication. (See the
description of smileys in section 14.5.)
Table
The header row of the table contains (from left to right):
The RT selector (EDT, T20 or T30), which determines which RT to display in the table
rows below. It is also linked to the main Spectrum selector and main Decay selector
The Frequency selector (which includes decrement and increment buttons),
determines the frequency of the readouts in the table rows below. The frequency selector is
linked to the Spectrum cursor and the frequency of the selected Decay
Readout of the RT value for the current measurement, before it is saved to a position
Each row of the table contains (from left to right):
A checkmark controlling whether the position is included or excluded from the Room
Average. Tap it to include (check) or exclude (uncheck) the position. All positions are by
default included in the Room Average
The position. Tap it to get a drop-down with two options, Select and View Annotations. Use
Select to select the position and View Annotations to view the list of annotations for the
position. You can add annotations on the position from this view – as described in Chapter 6
Possible annotation indicated by . Tap it to view the list of annotations for the position
The readout of the RT parameter at the frequency determined by the RT selector and
Frequency selector in the table header row. There might be a smiley to the right of the
readout warning about the quality of the readout. Tap the smiley to get more detailed
information about the warning
Vista de pagina 167