Hand-held Analyzer Types 2250 and 2270 – User Manual
9.2.3 Downgrade to an Earlier Version
It is possible to downgrade to an earlier version of the software:
1) Change the “latest” part of the server address (see Fig.9.2) to the requested software
package, for example, “SW22”
2) Press Check for updates.
3) Follow the procedure described in section 9.2.1 or section 9.2.2.
Note: Only software packages equal to or above SW24 can downgrade to another package.
Software packages down to SW21 are available for update through Internet.
9.3 How to Move a License
If you have more than one analyzer, you may want to share application software between the
analyzers. You can do this by moving the license from one analyzer to another by using
Measurement Partner Suite BZ-5503, together with the License Mover VP-0647.
If you lend out your analyzer, you may want to temporarily ‘un-install’ applications not needed.
This can be achieved by moving the license of the application to License Mover VP-0647. When
needed again, you move the license back to your analyzer.
Please consult the on-line help included with the BZ-5503 software for instructions on how to
move a license.
9.4 Troubleshooting
9.4.1 Measurements
If your measurement seems to be wrong, then:
• Check the cabling, if any
• Check that the microphone, including preamplifier, is correctly mounted in the top socket
(or correctly connected to the extension cable)
• Check that the Input parameter is set to Top Socket/Rear Socket in agreement with how
you are going to use the input. This is found by tapping the Main Menu icon , then
Setup, followed by Input
• Check that the transducer you have mounted on your analyzer is selected as the
Transducer parameter, this is also found on the Setup screen
• Check whether the parameters for the selected transducer (Transducer) are set correctly,
especially the Microphone Type and Polarization Voltage parameters. These are found by
tapping the Main Menu icon , then Transducers
• Check that the Sound Field and Windscreen Correction parameters have been set correctly
These are found by tapping the Main Menu icon , then Setup, followed by Input
• Check if the calibration is OK (make a new calibration using an external calibrator)