Hand-held Analyzer 2250 Light – User Manual
Manual Calibration
If no calibrator is available (or a known amplification is introduced, e.g., by analysing a tape
recorded signal) then you can type the sensitivity directly into the Sensitivity field. The
analyzer will be regarded as uncalibrated and the text ‘Uncal.’ will appear in the status field.
Charge Injection Calibration
You can check the complete measurement chain during long term measurement using Charge
Injection Calibration. See section 5.7.
5.3 Standard Calibration
To perform the acoustic calibration use Sound Calibrator Type 4231. It provides a stable sound
pressure at 1 kHz and has minimal susceptibility to environmental factors. The procedure itself is
relatively simple, and on the analyzer the procedure is referred to as the Standard Calibration
procedure.
The procedure for performing an acoustic calibration and instructions on how to fit the
calibrator are given below:
1) Stand away from loud sound sources that may interfere with the calibrators signal.
2) Switch on the analyzer: push .
3) Tap the Main Menu icon then Calibration to open the Calibration screen (Fig.5.1).
Fig.5.1
Calibration screen
This screen contains a bar graph showing the actual sound pressure level and three
placeholders for displaying information about the last calibration.
4) Following the first part of the instruction in the Status Field, fit Sound Calibrator
Type 4231 carefully onto the microphone of the analyzer. (To avoid handling vibrations