APPENDIX A
Setup Parameters
141
A.3 Statistics
Spectrum
*
X
C
Z
The frequency analysis (1/1-octave or 1/3-octave)
will be frequency weighted in accordance with this
parameter
Note: X = frequency weighting A or B. ‘A’ requires
that the Broadband (excl. Peak) parameter is set to
AC or AZ. ‘B’ requires that the Broadband (excl.
Peak) parameter is set to BC or BZ
Bandwidth
†
1/1-octave
1/3-octave
Bandwidth of frequency analysis
Bottom F. for Special L
eq
‡
,
**
12.5 Hz to
Top
Frequency
1/1-octave: 16 Hz – 8 kHz
1/3-octave: 12.5 Hz – 16 kHz
Top Freq. for Special L
eq
‡
,
††
Bottom
Frequency to
20 kHz
1/1-octave: 16 Hz – 8kHz
1/3-octave: 12.5 Hz – 16 kHz
*. Requires 1/1-octave or 1/3-octave Frequency Analysis Software BZ-7131 or BZ-7132 respectively
†. The Bandwidth parameter cannot be set in Type 2250-L – the setting is determined by the template.
‡. For BZ-7131, BZ-7132 and BZ-7133 only.
**. Bottom F. for Special Leq is f1 in the parameter LXeq(f1-f2)
††. Top F. for Special Leq is f2 in the parameter LXeq(f1-f2)
Table A.4 Statistics parameters
Parameter Values Comment
Broadband Statistics based on L
Xeq
L
XF
L
XS
The broadband statistics are based on sampling
the broadband parameter L
XF
or L
XS
each 10 ms
or L
Xeq
each second.
Note: X = frequency weighting A or B. ‘A’ requires
that the Broadband (excl. Peak) parameter is set
to AC or AZ. ‘B’ requires that the Broadband (excl.
Peak) parameter is set to BC or BZ
Percentile N1 0.1 to 99.9 User-defined percentile level where the value of
L
XN1
is exceeded for N1% of the elapsed time
Percentile N2 0.1 to 99.9 User-defined percentile level where the value of
L
XN2
is exceeded for N2% of the elapsed time
Percentile N3 0.1 to 99.9 User-defined percentile level where the value of
L
XN3
is exceeded for N3% of the elapsed time
Percentile N4 0.1 to 99.9 User-defined percentile level where the value of
L
XN4
is exceeded for N4% of the elapsed time
Table A.3 (Cont.) Frequency Settings parameters
Parameter Values Comment