B&K 2250 Light Manual de usuario Pagina 110

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 212
  • Tabla de contenidos
  • SOLUCIÓN DE PROBLEMAS
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 109
102
Hand-held Analyzer 2250 Light – User Manual
View whether the display of the profile during the measurement is frozen or not. When
is “animated” the display is being updated with new logged data during the measure-
ment. You can freeze the display update by tapping on the icon. This also freezes the icon.
Tap on the icon again to unfreeze the display
Step forwards or backwards through the logging intervals on all displays, using and
. (The icons are also connected to the profile cursor, so that any corresponding move-
ment backwards or forwards through the intervals in one display will be reproduced in the
other)
Some interaction with the display will automatically freeze and unfreeze the display (only
while measuring):
Tap the stylus on the profile. This will set the profile cursor and freeze the display update
of the profile. You can move the cursor to any point in the profile by using the left and
right arrow pushbuttons. The profile will be scrolled automatically, if necessary. Unfreeze
by tapping on
Tap and hold the stylus on the profile display and drag it to the left or right. This will
freeze the display and show two cursors. Next, remove the stylus from the display and a
drop-down menu appears with the possibility of setting or editing a marker or annotation
(see below). Select the required function from the drop-down menu. When the function
has been carried out, the profile unfreezes automatically and progresses as it did before
you tapped on the screen
The logged data are displayed as two profiles. The Profile Overview with the entire profile and
the Profile with 100 logging samples.
The Profile Overview
An overview of the entire profile is displayed across the screen on the topmost part of the graph
area.
The overview is based on L
Xeq,1s
(X = A or B depending on the Broadband (excl. Peak)
Frequency Settings parameter). The Y-axis is auto-zoomed.
When you have logged for more than 4 minutes, then each pixel on the X-axis will cover more
than 1 s. The overview will then display from the minimum L
Xeq,1s
to the maximum L
Xeq,1s
within the interval covered by the pixels on the X-axis.
Tap in the Profile Overview to select the part of the entire profile to display in the Profile
below.
The Profile
The Profile displays a hundred samples of the logged data.
Select which parameter to view by tapping the parameter field in the line above the profile.
Markers are displayed between the profile and the profile parameter. Marker 1 (Exclude) is the
topmost positioned marker. The name of the marker is displayed if the marker overlaps the
cursor position. If a signal has been recorded, then a Sound marker (Marker 6) is displayed
above the profile – the marker covers the exact time of the signal recording.
Vista de pagina 109
1 2 ... 105 106 107 108 109 110 111 112 113 114 115 ... 211 212

Comentarios a estos manuales