Display
Action
Remark
Conductivity and temperature
are displayed.
Remove the sensor from the
calibration solution and clean it.
Place sensor in process.
The measured value is shown in
the main display alternately with
“Hold”; “enter” flashes.
End calibration with
enter.
Security prompt.
After end of calibra-
tion, the outputs
remain in Hold
mode for approx.
20 sec.