B&K Reference 7250 Series II Manual de instrucciones

Busca en linea o descarga Manual de instrucciones para Amplificadores de audio B&K Reference 7250 Series II. B&K Reference 7250 Series II Instruction manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 108
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
Instruction manual
Transmitter Cond Ind 7100e/2(X)H
70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 1
Vista de pagina 0
1 2 3 4 5 6 ... 107 108

Indice de contenidos

Pagina 1 - Instruction manual

Instruction manualTransmitter Cond Ind 7100e/2(X)H70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 1

Pagina 2

10Transmitter Cond Ind 7100e/2(X)EC-Type-Examination Certificate70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 10

Pagina 3 - Contents

100Transmitter Cond Ind 7100e/2(X)70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 100

Pagina 4

22 mA signal for error message . . . . . . . . . . . . . . . . . . . . 51AAccessories . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 5 - Safety information

Configuration steps . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 6

HHART communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . 31Hold mode . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 7

PPackage contents . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16Panel-mount kit . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 8 - EC Declaration of Conformity

UUser interface . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26WWarranty . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 9

70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 106

Pagina 10

70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 107

Pagina 11

TA-194.301-MTE02 011005Software version: 1.1Mettler-Toledo GmbH, Process AnalyticsIndustrie Nord, CH-8902 Urdorf, SwitzerlandPhone + 41 44 736 22 11,

Pagina 12

1170098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 11

Pagina 13

12Transmitter Cond Ind 7100e/2(X)70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 12

Pagina 14

131. Supplement to EC-Type-Examination Certificate70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 13

Pagina 15 - Overview of Transmitter

14Transmitter Cond Ind 7100e/2(X)H70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 14

Pagina 16 - Assembly

15Overview of Transmitter12345678Inputreceive hireceive losend losend hishieldnot connected(7100e/2H)equipotentialbonding(7100e/2XH)RTDRTDTempinput101

Pagina 17 - Fig.: Mounting plan

16Transmitter Cond Ind 7100e/2(X)Package contentsCheck the shipment for transport damage and completeness.The package should contain:Assembly11109876

Pagina 18 - Pipe mounting, panel mounting

17Mounting plan1441441542848032214310527726,212341 Cable gland (3 pieces)2 Breakthroughs for cable glandor conduit 1/2”,dia 21.5 mm (2 breakthroughs)C

Pagina 19 - Fig.: Panel-mount kit

18Transmitter Cond Ind 7100e/2(X)40 6013212345Fig.: Pipe-mount kit1 Protective hood (if required)2 Hose clamps with worm gear drive to DIN 3017 (2 pie

Pagina 20 - Division 2 wiring

1912345max. 2578 271...221 Screws (4 pieces)2 Gasket (1 piece)3 Panel4 Span pieces (4 pieces)5 Threaded sleeves (4 pieces)Fig.: Panel-mount kitPanel c

Pagina 21 - Terminal assignments

WarrantyDefects occurring within 1 years from delivery date shall be remedied freeof charge at our plant (carriage and insurance paid by sender).Subje

Pagina 22

20Transmitter Cond Ind 7100e/2(X)Installation and connectionInformation on installation• Installation may only be carried out by trained experts in ac

Pagina 23

21Terminal assignmentsFig.: Terminal assignments Transmitter Cond Ind 7100 e/2HFig.: Terminal assignments Transmitter Cond Ind 7100 e/2XH70098_E_CI710

Pagina 24 - Wiring examples

22Transmitter Cond Ind 7100e/2(X)H1 Recommended strippinglengths for multi-core cables2 Pulling out the terminalsusing a screwdriver(also see 6)3 Cabl

Pagina 25 - Cond Ind 7100e/2(X)H

2370098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 23

Pagina 26 - User interface and display

Transmitter Cond Ind 7100e/2(X)24Wiring examplesInPro 7250 ST sensorreceive hireceive loshieldRTDcoreRDBNBUshield (GN/YE)WHcoax BKGNsend losend hiCond

Pagina 27

25InPro 7250 HT sensorreceive hireceive loshieldRTDcoreYEVTcoreshield (GN/YE)WHcoax RDcoax BKGNsend losend hiCond Ind 7100e/2(X)H GYRTD70098_E_CI7100_

Pagina 28 - Operation: Keypad

26Transmitter Cond Ind 7100e/2(X)User interface and display1 Display2 Mode indicators (no keys), from left to right:- Measuring mode- Calibration mode

Pagina 29 - Safety functions

271234 56 78910111213161720181915141 Passcode entry2 Display of measured variable*3 Temperature4 Current output5 Limit values6 Alarm7 Sensocheck8 Cali

Pagina 30

28Transmitter Cond Ind 7100e/2(X)Operation: KeypadStart, end calibrationStart, end configurationSelect digit position(selected position flashes)Edit d

Pagina 31

29Safety functionsSensocheck, Sensoface sensor monitoringSensocheck continuously monitors the sensor and lines. The primary and its lines are monitore

Pagina 32

Safety information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .5Intended use . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Pagina 33 - Passcodes (factory setting)

30Transmitter Cond Ind 7100e/2(X)Safety functionsHold modeDisplay: The Hold mode is a safety state during configuration and cali-bration. The loop cur

Pagina 34 - Configuration

31OutputsCurrent output / Loop currentThe loop current is controlled by the process variable selectedin the configuration.The current start and end ca

Pagina 35 - Menu group

32Transmitter Cond Ind 7100e/2(X)H70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 32

Pagina 36 - Selection / Default

33Passcodes (factory setting)The passcodes allow fast access to the functionsCalibrationKey+passcodeDescription Cal Info00001100Product calibration110

Pagina 37 - Individual settings

34Transmitter Cond Ind 7100e/2(X)ConfigurationIn the Configuration mode you set the device parameters.Activate with confEnter passcode “1200”*. Edit p

Pagina 38

35Menu groupCodeCurrent outputo1.Temperature compensationtc.AlarmsettingsAL.DisplaySelect menu groupPrevious menugroup:Menu structure of configuration

Pagina 39 - After correct input a

36Transmitter Cond Ind 7100e/2(X)Overview of configuration steps7250 IPR, OthermS/cm, S/m, SAL, %LIN / LOGtc.UnITSelect temperature unit°C / °FSelect

Pagina 40

Code Parameter Default Individualsettings settingso1.SnSR Sensor selection 7250 IPR ____________- With “Other” selected:- Cell factor 2.175

Pagina 41 - Code Display

38Transmitter Cond Ind 7100e/2(X)ConfigurationCurrent output: Select sensor type.Select sensorSelect measured variableEnter current startEnter current

Pagina 42

7250 IPR(Other) 39Note::Characters represented in gray are flashing and can be edited.Code DisplayActionChoiceso1.Select configuration(Press conf.)Ent

Pagina 43 - Concentration measurement

Transmitter Cond Ind 7100e/2(X)ContentsTemperature compensation . . . . . . . . . . . . . . . . . . . . . . . . . . . .52Alarm settings . . . . . . .

Pagina 44

40Transmitter Cond Ind 7100e/2(X)ConfigurationCurrent output: Select measured variable.Select sensorSelect measured variableEnter current startEnter c

Pagina 45

41Code DisplayActionChoiceso1.Select measured variable: Select with  arrow keyProceed with enterConductivity:• 0.000 ... 9.999 mS/cm• 00.00 ... 99.99

Pagina 46

42Transmitter Cond Ind 7100e/2(X)ConfigurationCurrent outputConcentration measurement: Select process solution.Select sensorSelect measured variableEn

Pagina 47

43Code DisplayActionChoiceso1.Only with 00.00 % can you selectthe process solution:Select witharrow key NaCl* -01-HCl* -02--07-NaOH* -03--10-H2SO4* -

Pagina 48

44Transmitter Cond Ind 7100e/2(X)ConfigurationOutput current. LIN characteristic. Current start/endSelect sensorSelect measured variableEnter current

Pagina 49 - Conductivity

45CodeLIN(LIN / LOG)DisplayActionChoiceso1.Select output characteristicSelect with  arrow keyProceed with enter(Step omitted for % (Conc) orSAL)000.0

Pagina 50

46Transmitter Cond Ind 7100e/2(X)ConfigurationOutput current. LOG characteristic. Current start / endSelect sensorSelect measured variableEnter curren

Pagina 51

47CodeDisplayActionChoiceso1.With LOG selectednot with % (Conc) or SAL):• Enter lower end of scale(=current start)Select with  key, edit numberwith 

Pagina 52

48Transmitter Cond Ind 7100e/2(X)ConfigurationOutput. Time constant of output filterSelect sensorSelect measured variableEnter current startEnter curr

Pagina 53 - Proceed with enter

490 sec0 ... 120 secCodeDisplayActionChoiceso1.Time constant of output filterDefault setting: 0 s (inactive).To specify a time constant: Selectwith 

Pagina 54

5Safety informationBe sure to read and observe the following instruc-tions!The device has been designed using state of the art technolo-gies and it co

Pagina 55 - (xxxx s)

50Transmitter Cond Ind 7100e/2(X)ConfigurationOutput. Output current during Error and HOLD.Select sensorSelect measured variableEnter current startEnt

Pagina 56 - Display Action

51OFF (OFF / ON)CodeDisplayActionChoiceso1.22 mA signal for error messageSelect with  arrow keyProceed with enterOutput signal for HOLDLAST: During H

Pagina 57

52Transmitter Cond Ind 7100e/2(X)ConfigurationTemperature compensation1. Press conf key.2. Enter passcode 1200*.3. Select Temperature compensationmenu

Pagina 58 - Calibration

53OFF(OFFLINnLF)CodeDisplayActionChoicestc.Only with linear temperature com-pensation (LIN) selected:Enter temperature coefficient*.Select position wi

Pagina 59

54Transmitter Cond Ind 7100e/2(X)ConfigurationAlarm settings1. Press conf key.2. Enter passcode 1200*.3. Select Alarm settings menu groupusing arrow k

Pagina 60 - Display

55CodeDisplayActionChoicesOFF (ON / OFF)AL.Select Sensocheck(Continuous monitoring of sensorproperties)Select with  key.Proceed with enterAlarm delay

Pagina 61 - Display Action

56Transmitter Cond Ind 7100e/2(X)Passcodes according to FDA 21 CFR Part 11Access to the device functions can be protected with adjustablepasscodes. To

Pagina 62

57Caution!If you have lost theAdministrator pass-code, the PasscodeEditor cannot becalled up! Please con-sult our technicalsupport!Display ActionRemar

Pagina 63

58Transmitter Cond Ind 7100e/2(X)CalibrationCalibration adjusts the device to the sensor.Activate with calEnter passcode*: • Entry of cell factor 1100

Pagina 64 - Product calibration

59Information on calibrationCalibration can be performed by:• Entry of cell factor• Determining the cell factor with a known calibration solution taki

Pagina 65

6Transmitter Cond Ind 7100e/2(X)Safety precautions for installation• The stipulations of EN 60079-10 / EN 60079-14 must be observed during commissioni

Pagina 66 - Zero calibration in air

60Transmitter Cond Ind 7100e/2(X)Display ActionRemark Enter cell factor:Select with  key, edit number with  key.Conductivity and temperatureare alte

Pagina 67

61Display Action Remark Conductivity and temperatureare displayed.The measured value is shown inthe main display alternately with“Hold”; “enter” flas

Pagina 68

62Transmitter Cond Ind 7100e/2(X)Display Action Remark Calibration with calibration solutionPress cal key, enter code 0110*Select with key, edit numb

Pagina 69

63Display Action Remark The determined cell factor andzero point are displayed.Confirm cell factor with enter.Clean sensor and re-place it inthe proc

Pagina 70 - Temperature probe adjustment

64Transmitter Cond Ind 7100e/2(X)The measurement process is only interrupted briefly. During product calibration the sensor remains in the process.Cal

Pagina 71 - Diagnostics functions

Product calibration step 2:When the sample value hasbeen determined, call up theproduct calibration once more(Press cal key. Enter passcode1105*.65Dis

Pagina 72

66Transmitter Cond Ind 7100e/2(X)DisplayAction Remark When there has notbeen an entry for6 sec, the lowerdisplay alternatelyshows the conductiv-ity an

Pagina 73 - Operating states

67Display Action Remark The cell factor and zero pointare displayed.Press enter to confirm thecalibration data. Place sensor in process.The measured v

Pagina 74 - Error messages (Error Codes)

68Transmitter Cond Ind 7100e/2(X)DisplayAction Remark When there has notbeen an entry for6 sec, the lowerdisplay alternatelyshows the conductiv-ity an

Pagina 75 - Display of last error message

69DisplayAction Remark Conductivity and temperatureare displayed.Remove the sensor from thecalibration solution and clean it.Place sensor in process.T

Pagina 76 - Sensoface

7Intended useThe Transmitter Cond Ind 7100e/2(X) is used for measurement ofelectrical conductivity and temperature in liquids in the field ofbiotechno

Pagina 77 - Problem

70Transmitter Cond Ind 7100e/2(X)Display Remark MeasurementIn the measuring mode the main display shows theconfigured process variable (conductivity,

Pagina 78

71Display of the output currentPress enter while in measuring mode.For 5 sec, the secondary display shows the outputcurrent instead of the temperature

Pagina 79 - Appendix

72Transmitter Cond Ind 7100e/2(X)Specify output currentfor testing the connected peripheralsPress conf while in measuring mode and enter pass-code 555

Pagina 80 - Specifications

Operating state73MeasurementOperating statesCal Info(cal) 0000*20 secError Info(conf) 0000*20 secCalibration(cal) 1100*Temp adjustment(cal) 1015*LEDTi

Pagina 81

Measuredvalueflashes74Transmitter Cond Ind 7100e/2(X)Red LEDOut 1 (22 mA)Error messages (Error Codes)Sensor• Measurement range violation• SAL > 45

Pagina 82

75Sensocheck:Primary coilSensocheck:Secondary coilRed LEDOut 1 (22 mA)Current outputCurrent below 3.8 mAErrorIcon(flashes)ProblemPossible causesERR 11

Pagina 83

76Transmitter Cond Ind 7100e/2(X)SensocheckContinuously monitors the primary coil and its lines for shortcircuits and the secondary coil and its lines

Pagina 84

77StatusDisplayProblem Sensor defectShort circuit in primary coilOpen circuit in secondary coil (see also Error messagesErr 33 and Err 34, Pg 75).Temp

Pagina 85

78Transmitter Cond Ind 7100e/2(X)70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 78

Pagina 86 - Calibration solutions

79AppendixProduct line and accessoriesDevices Order no.Transmitter Cond Ind 7100e/2H 52 121 257Transmitter Cond Ind 7100e/2XH 52 121 258Mounting acces

Pagina 87 - Sodium chloride solutions

8Transmitter Cond Ind 7100e/2(X)EC Declaration of Conformity70098_E_CI7100_2XH.QXD 27.07.2005 9:27 Uhr Seite 8Declaration of conformity Konformität

Pagina 88

80Transmitter Cond Ind 7100e/2(X)HConductivity input Input for electrodeless conductivity sensorsEffective range Conductivity 0.000 mS/cm ... 1999 mS/

Pagina 89 - Concentration curves

81Sensor standardizationOperating modes • Entry of cell factor with simultaneous display of conductivity and temperature • Input of conductivity of ca

Pagina 90

Transmitter Cond Ind 7100e/2(X)Loop current 4 ... 20 mA floating Supply voltage 14 ... 30 VMeasured variable * Conductivity, concentration, or salinit

Pagina 91

83Service functionsCurrent source Loop current specifiable 3.8 ... 22.00 mADevice self-test Automatic memory test (RAM, FLASH, EEPROM)Display test Di

Pagina 92

84Transmitter Cond Ind 7100e/2(X)HEnclosure Molded enclosure made of PBT(polybutylene terephtalate)Color Bluish gray RAL 7031Mounting • Wall mounting

Pagina 93 - -05- Nitric acid HNO

8570098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 85

Pagina 94 - FM Control Drawing

Transmitter Cond Ind 7100e/2(X)86Calibration solutionsPotassium chloride solutions(Conductivity in mS/cm)Temperature Concentration *[°C] 0.01 mol/l 0.

Pagina 95

87Sodium chloride solutions(Conductivity in mS/cm)0123456789101112131415161718192021222324252627282930313233343536134.5138.6142.7146.9151.2155.5159.91

Pagina 96 - Explosion protection

88Transmitter Cond Ind 7100e/2(X)Concentration measurementMeasurement rangesFor the solutions listed above, the transmitter can determine the sub-stan

Pagina 97

89Conductivity in dependence on substance concentration and processtemperature for sodium chloride (NaCl)Concentration curves-01- Sodium chloride solu

Pagina 98 - CSA Control Drawing

9Conformity with FDA 21 CFR Part 11In their directive “Title 21 Code of Federal Regulations, 21 CFRPart 11, Electronic Records; Electronic Signatures”

Pagina 99

90Transmitter Cond Ind 7100e/2(X)Conductivity in dependence on substance concentration and processtemperature for hydrochloric acid (HCl)Source: Haase

Pagina 100

91Conductivity in dependence on substance concentration andprocess temperature for sodium hydroxide solution (NaOH)-03- Sodium hydroxide solution NaOH

Pagina 101

92Transmitter Cond Ind 7100e/2(X)Conductivity in dependence on substance concentration and processtemperature for sulfuric acid (H2SO4), Source: Darli

Pagina 102

93Conductivity in dependence on substance concentration and processtemperature for nitric acid (HN03)Source: Haase/Sauermann/Dücker; Z. phys. Chem. Ne

Pagina 103

94Transmitter Cond Ind 7100e/2(X)HFM Control Drawing70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 94

Pagina 104

9570098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 95

Pagina 105

96Transmitter Cond Ind 7100e/2(X)HExplosion protection70098_E_CI7100_2XH.QXD 27.07.2005 9:28 Uhr Seite 96

Pagina 106

97Warnings and notes to ensure safe operationWarning: Do not disconnect equipment unless power hasbeen switched off.Warning: Clean only with antistati

Pagina 107

98Transmitter Cond Ind 7100e/2(X)HCSA Control DrawingConductivity Measuring LoopTerminals1, 2, 3, 4, 5, 6Uo, Vsc Io, Isc Po Co, Ca Lo, LaIIC (GRP A, B