Transmitter Cond Ind 7100e/2(X)
Display
Action
Remark
Calibration with calibration solution
Press cal key, enter code 0110*
Select with
key, edit number
with
key, proceed with enter.
Ready for calibration
Dismount and clean sensor
If an invalid code is
entered, the trans-
mitter returns to
measuring mode.
Display (3 sec)
Transmitter in Hold
mode, measured
value frozen.
Sensoface inactive.
Be sure to use known calibration solutions and the respective tempera-
ture-corrected conductivity values (see Calibration solutions Pg86).
During the calibration procedure the temperature should be kept con-
stant.
Immerse sensor in calibration
solution.
Enter the temperature-corrected
conductivity value of the
calibration solution:
Select with
key, edit number
with
key.
Cell factor and temperature are
alternately displayed in the lower
display.
Confirm entry with enter.
When there has not
been an entry for
6 sec, the lower
display alternately
shows the conduc-
tivity and tempera-
ture value.
* Factory setting