GPS Annotations
A text annotation with GPS information can be attached (Latitude,
Longitude, Altitude and position error). Requires connection to a GPS
receiver
Data Management
PROJECT TEMPLATE
Defines the display and measurement settings. Setups can be locked
and password protected
PROJECT
Measurement data stored with the project template
JOB
Projects are organised in jobs
Explorer facilities for easy management of data (copy, cut, paste, delete,
rename, view data, open project, create job, set default project name)
Software Specifications – 1/1-octave Frequency Analysis Software for 2250 Light BZ-7131 and 1/3-octave
Frequency Analysis Software for 2250 Light BZ-7132
The specifications for BZ-7131 and BZ-7132 include the specifications
for Sound Level Meter Software for 2250 Light BZ-7130. BZ-7131 and
BZ-7132 add:
Standards
Conforms with the following National and International Standards:
IEC 61260 (199507) plus Amendment 1 (200109), 1/1-octave Bands
and 1/3-octave Bands, Class 0
ANSI S1.111986, 1/1-octave Bands and 1/3-octave Bands, Order 3,
Type 0 C
ANSI S1.11 2004, 1/1-octave Bands and 1/3-octave Bands, Class 0
Frequency Analysis
CENTRE FREQUENCIES
1/1-octave Band Centre Frequencies (BZ-7131 only):
16 Hz to 8 kHz
1/3-octave Band Centre Frequencies (BZ-7132 only):
12.5 Hz to 16 kHz
MEASUREMENTS
X = frequency weightings A, B, C or Z
Spectra for Display and Storage:
Spectra for Display Only:
Single Values
MEASURING RANGES
When using Microphone Type 4950:
BZ-7131, 1/1-octave:
Dynamic Range: From typical noise floor to max. level for a pure tone
signal at 1 kHz: 5.9 to 140 dB
Linear Operating Range: In accordance with IEC 61260: 22.9 dB to
140 dB
BZ-7132, 1/3-octave:
Dynamic Range: From typical noise floor to max. level for a pure tone
signal at 1 kHz: 1.0 to 140 dB
Linear Operating Range: In accordance with IEC 61260: 20.1 dB to
140 dB
MEASUREMENT DISPLAYS
Spectrum: One or two spectra superimposed + A/B and C/Z broadband
bars
Table: One or two spectra in tabular form
Y-axis: Range: 5, 10, 20, 40, 60, 80, 100, 120, 140 or 160 dB. Auto
zoom or auto scale available
Cursor: Readout of selected band
Software Specifications – Logging Software for 2250 Light BZ-7133
The specifications for BZ-7133 include the specifications for Sound Level
Meter Software for 2250 Light BZ-7130. BZ-7133 adds:
Logging
MEASUREMENTS
Measurement data logged at pre-set periods into files on:
SD Card
USB Memory Stick
Logging Period: From 1 s to 24 hr with 1 s resolution
Fast Logging: L
AF
and L
Aeq
can be logged every 100 ms, irrespective
of logging period
Broadband Data Stored at each Logging Interval: All, or up to 10
selectable broadband data including weather data and L
Aeq,T,mov
Broadband Statistics Stored at each Logging Interval: Full
distribution, or none
Spectrum Data Stored at each Logging Interval: All, or up to three
selectable spectra (license for BZ-7131 or BZ-7132 required)
Logging Time: From 1 s to 31 days with 1 s resolution
Measurement Total: For the logging time, in parallel with logging: All
broadband data, statistics and spectra (license for BZ-7131 or BZ-7132
required)
Automatic reboot and resume of operation in case of power failure
MEASUREMENT DISPLAYS
Profile: Graphical display of selectable measurement data versus time.
Fast display of next or previous marker; Profile overview of entire
measurement
Y-axis: Range: 5, 10, 20, 40, 60, 80, 100, 120, 140 or 160 dB. Auto
zoom or auto scale available
X-axis: Scroll facilities
Cursor: Readout of measurement data at selected time
MARKERS
Five user-definable markers for on-line marking of noise sources or
events anywhere in the profile.
Markers are set using the stylus on the touch screen, or the three marker
pushbuttons
L
Xeq
L
XSmax
L
XFmax
L
XSmin
L
XFmin
L
XS
L
XF
SIL PSIL SIL3
L
Xeq
(f1-f2)
a
a
where f1 and f2 are frequency bands in the spectrum.
NR NR Decisive Band
RC RC Classification
NCB NCB Classification
NC NC Decisive Band
Loudness Loudness Level (BZ-7132 only)